In vivo dosimetry with alanine/electron spin resonance dosimetry to evaluate the urethra dose during high-dose-rate brachytherapy

alertIcon.gif

Publication date: Available online 9 May 2017
Source:Brachytherapy
Author(s): Daniela Wagner, Markus Hermann, Andrea Hille
PurposeTo determine the dose of the urethra during Ir-192 high-dose-rate brachytherapy in 15 patients, a new detector form developed for in vivo measurements was used as described by Anton et al. 2009.Methods and MaterialsThe detector consists of alanine powder with paraffin as binder which was filled in a shrinkable tube for measurements. The alanine strand so produced was inserted in the foley catheter during high-dose-rate brachytherapy of the prostate. The measured dose was compared with the dose calculated by the treatment planning system SWIFT for 15 patients. After 8 patients, a marker was introduced for better positioning of the alanine strand using transrectal ultrasound images.ResultsThe measurements of the dose of the urethra agreed very well with the dose calculated by SWIFT within the estimated standard uncertainty of the method for 8 patients with a mean absolute deviation of 0.08 Gy. However, for the other 7 patients, a mean absolute deviation between delivered and measured dose of −5.13 Gy was seen. For these patients, the active volume of the alanine dosimeter could not properly be reconstructed on the ultrasound images.ConclusionThe method presented in this study is useful for quality control of irradiations in vivo. To reconstruct the active volume on the corresponding ultrasound images, the correct application of the alanine strand in the urethra is very important. This procedure needs a well-trained physician.



http://ift.tt/2pwCmMj

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου