DEVELOPMENT OF AN IONIZATION CHAMBER FOR LOW AND MEDIUM ENERGY PHOTON DOSIMETRY.

Related Articles

DEVELOPMENT OF AN IONIZATION CHAMBER FOR LOW AND MEDIUM ENERGY PHOTON DOSIMETRY.

Radiat Prot Dosimetry. 2018 Apr 05;:

Authors: Singh SK, Kulkarni MS

Abstract
Measurement of dose due to low and medium energy X-ray using an ionization chamber needs special considerations as the wall thickness of the detector plays an important role in signal generation. A thin and plane wall ionization chamber having 900 cc volume was developed to study its calibration coefficient with air kerma (Kair) rate and ambient dose equivalent (H*(10)) rate at various X-ray beam qualities generated from a dosimetry grade X-ray machine. Optimized ionization chamber wall thickness was determined through measurements where a flat energy response of the ionization chamber could be established. The measurement shows that the air kerma rate based calibration coefficient for ionization chamber varies up to 45% and ambient dose equivalent rate based calibration coefficient varies up to 20% for wall thickness 1.2-1200 mg/cm2 in the energy range 17.6-213 keV.

PMID: 29660104 [PubMed - as supplied by publisher]