Mechanical properties of different esthetic and conventional orthodontic wires in bending tests

Abstract

Aims

The goal of this study was to determine the mechanical properties of different esthetic and conventional orthodontic wires in three-point and four-point bending tests, and in a biomechanical test employing three bracket systems.

Methods

The behavior of round wires with a diameter of 0.46 mm (0.018″) were investigated: uncoated nickel titanium (NiTi) wires, surface modified NiTi wires; FLI® Orthonol Wire® and glass fiber reinforced plastic wires. The biomechanical bending test was performed using the following bracket types: metal brackets (Discovery®, Dentaurum), ceramic brackets (Fascination®, Dentaurum), and plastic brackets (Elegance®, Dentaurum). All bending tests were performed in the orthodontic measurement and simulation system (OMSS) at a temperature of 37 °C. The classical three-point bending test was performed according to an ISO standard (DIN EN ISO 15841:2007) using the appropriate thrust die and supports with a predefined span of 10 mm. In the other tests the supports or interbracket distances were chosen such that the free wire length was also 10 mm (5 mm between adjacent brackets). All wires were loaded centrally to a maximum of 3.1 and 3.3 mm in the biomechanical test, respectively. The force was measured upon unloading with a loading velocity of 1 mm/min. Each specimen was loaded twice and a total of 10 specimens tested for each product. Weighted means and the error of the weighted mean were calculated for each product.

Results

Fiber reinforced wires displayed lowest forces in three-point bending with values of 0.4 N at a displacement of 1 mm and 0.7 N at a 2 mm displacement. In four-point bending the forces were 0.9 N and 1.4 N, respectively, at the same displacements. Almost all of the translucent wires showed fracture upon bending at displacements greater than 3 mm, independent of the bending test and bracket type. The different investigated NiTi wires, surface modified or conventional, only showed minor variation, e.g., 2.2 N for rematitan® Lite White and 2.0 N for rematitan®, 2.1 N for FLI® Coated Orthonol® and 1.7 N for Orthonol® in four-point bending. The rhodinized wire generated forces between these values (2.1 N).

Conclusion

The translucent wires had the lowest forces in all three bending tests; however, displacements above 3 mm resulted in increased risk of fracture. Forces of investigated NiTi wires were very high and in part above clinically recommended values.



http://ift.tt/2h5UyvQ