Σφακιανάκης Αλέξανδρος
ΩτοΡινοΛαρυγγολόγος
Αναπαύσεως 5 Άγιος Νικόλαος
Κρήτη 72100
00302841026182
00306932607174
alsfakia@gmail.com

Αρχειοθήκη ιστολογίου

! # Ola via Alexandros G.Sfakianakis on Inoreader

Η λίστα ιστολογίων μου

Κυριακή 26 Μαρτίου 2017

COMPARING MEASURED AND CALCULATED DOSES IN INTERVENTIONAL CARDIOLOGY PROCEDURES.

Related Articles

COMPARING MEASURED AND CALCULATED DOSES IN INTERVENTIONAL CARDIOLOGY PROCEDURES.

Radiat Prot Dosimetry. 2017 Mar 16;:1-5

Authors: Oliveira da Silva MW, Canevaro LV, Hunt J, Rodrigues BB

Abstract
Interventional cardiology requires complex procedures and can result in high doses and dose rates to the patient and medical staff. The many variables that influence the dose to the patient and staff include the beam position and angle, beam size, kVp, filtration, kerma-area product and focus-skin distance. A number of studies using the Monte Carlo method have been undertaken to obtain prospective dose assessments. In this paper, detailed irradiation scenarios were simulated mathematically and the resulting dose estimates were compared with real measurements made previously under very similar irradiation conditions and geometries. The real measurements and the calculated doses were carried out using or simulating an interventional cardiology system with a flat monoplane detector installed in a dedicated room with an Alderson phantom placed on the procedure table. The X-ray spectra, beam angles, focus-skin distance, measured kerma-area product and filtration were simulated, and the real dose measurements and calculated doses were compared. It was shown that the Monte Carlo method was capable of reproducing the real dose measurements within acceptable levels of uncertainty.

PMID: 28338948 [PubMed - as supplied by publisher]



http://ift.tt/2nVdXTL

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου

Αρχειοθήκη ιστολογίου