Analysis of radiation doses and dose reduction strategies during cerebral digital subtraction angiography.

Related Articles

Analysis of radiation doses and dose reduction strategies during cerebral digital subtraction angiography.

World Neurosurg. 2017 Jan 12;:

Authors: Yi HJ, Sung JH, Lee DH, Kim SW, Lee SW

Abstract
OBJECTIVE: Adverse effects of increased use of cerebral digital subtraction angiography (DSA) include radiation-induced skin reactions and increased risk of malignancy. This study aimed to identify a method for reducing radiation exposure during routine cerebral DSA.
METHODS: A retrospective review of 138 consecutive adult patients who underwent DSA with a biplane angiography system (Artis Zee, Siemens, Germany) from September 2015 to February 2016 was performed. In January 2016, the 'dose parameter' was reset by the manufacturer from 2.4μGy to 1.2μGy. Pre-dose (Group 1) and post-dose parameter reduction (Group 2) groups were established. Angiograms and procedure examination protocols were reviewed according to patient age, gender, and diagnosis while angiography techniques were reviewed on the basis of the following radiation dose parameters: fluoroscopy time, reference point air kerma (Ka,r; in mGy), and kerma area product (PKA; in μGym(2)).
RESULTS: The mean Ka,r values in Groups 1 and 2 were 1841.5 mGy and 1274.8 mGy, respectively. The mean PKA values in Groups 1 and 2 were 23212.5 μGym(2) and 14854.0 μGym(2), respectively. Ka,r and PKA values were significantly lower in Group 2 compared with Group 1 (p<0.001). Among individual factors, young age was a determining factor for reduced fluoroscopy time (p<0.001), Ka,r (p=0.047), and PKA (p=0.022).
CONCLUSIONS: Increased awareness of radiation risks, as well as the establishment of strategies to reduce radiation dose, led to lower radiation doses for DSA. The use of appropriate examinations and low-dose parameters in fluoroscopy contributed significantly to the radiation dose reductions.

PMID: 28089806 [PubMed - as supplied by publisher]