Comparative Analysis of Serum Allergen sIgE Detected by Western Blotting and Fluorescent Enzyme Immunoassay.

Related Articles

Comparative Analysis of Serum Allergen sIgE Detected by Western Blotting and Fluorescent Enzyme Immunoassay.

Clin Lab. 2017 Mar 01;63(3):487-493

Authors: Chen X, Zheng P, Wei N, Huang H, Luo W, Lin Y, Sun B

Abstract
BACKGROUND: Various new methods to diagnose allergens have gradually developed in recent years. The present study aimed to evaluate the diagnostic performance and clinical utility of Allergy-Q detection system (Proteometech Inc., Korea) on serum allergen-specific IgE (sIgE).
METHODS: Immunoblotting (Allergy-Q system) and fluorescent enzyme immunoassay (ImmunoCAP system, Phadia Inc., Sweden) were used to detect serum total IgE and allergen sIgE of 244 cases of allergic diseases, followed by comparative analysis of the results. The study focused on analyzing 15 kinds of common allergens.
RESULTS: Comparison of the two methods showed that the diagnostic performance of qualitative results of AllergyQ system for the majority of allergens was satisfactory, with Dermatophagoides farinae having the highest agreement (96.4%) and sensitivity (97.8%) and Dermatophagoides pteronyssinus having the highest specificity (100%), positive predictive value (100%), and Youden's index (J = 0.951). Scatter plots and Spearman's correlation analysis showed correlations between the quantitative results of the two methods, with the highest correlation for Dermatophagoides pteronyssinus (r = 0.875, p < 0.001).
CONCLUSIONS: For most common allergens, the Allergy-Q system is an effective method for diagnosis. It is safe, convenient, economical, accurate, and efficient in screening multi-allergens. Moreover, it is good for semi-quantitative detection of allergens in the basic hospitals that are unable to carry out the ImmunoCAP detection, and also for patients with fewer samples. However, its clinical application needs to be validated using a number of samples from a variety of countries and regions.

PMID: 28271691 [PubMed - in process]