Σφακιανάκης Αλέξανδρος
ΩτοΡινοΛαρυγγολόγος
Αναπαύσεως 5 Άγιος Νικόλαος
Κρήτη 72100
00302841026182
00306932607174
alsfakia@gmail.com

Αρχειοθήκη ιστολογίου

! # Ola via Alexandros G.Sfakianakis on Inoreader

Η λίστα ιστολογίων μου

Τετάρτη 31 Μαΐου 2017

Optimal combination of anti-scatter grids and software correction for CBCT imaging.

Optimal combination of anti-scatter grids and software correction for CBCT imaging.

Med Phys. 2017 May 29;:

Authors: Stankovic U, Ploeger LS, van Herk M, Sonke JJ

Abstract
PURPOSE: CBCT has been widely adopted in clinical practice for image-guided radiotherapy. Soft tissue contrast and Hounsfield units are impaired to presence of scattered radiation. In our previous work we proposed a high selectivity anti-scatter grid (ASG) as a possible solution to the problem. An alternative approach is the application of iterative scatter correction using deconvolution with scatter point spread function (PSF). The purpose of this work was to compare the performance of ASGs with different selectivity with and without the iterative and uniform scatter corrections in terms of CBCT image quality. A secondary objective of this study was to develop a novel measurement approach to measure the scatter point spread functions.
METHODS: The scatter PSF was modeled as a sum of two bivariate Gaussian functions. The PSF parameters were estimated from a series of transmission measurements through polystyrene slabs of varying thickness with lead partial beam-blocker for three different anti-scatter grid designs ranging from low (5.6), medium (9) and high (11) selectivity. The scatter correction scheme is based on iterative convolution of the current estimate of the primary with the scatter PSF until the root mean square deviation of the measured projection and the sum of the estimate of primary and scatter falls below a pre-defined threshold. The image quality was evaluated with the CIRS CBCT Image Quality and Electron Density phantom in a head and neck and pelvis configuration and the CIRS Virtual Male Human Patient. The image quality was quantified by the contrast-to-noise ratio (CNR) relative to the uncorrected scans and the root mean square deviation (RMSD) of the average gray values for different regions with respect to the nominal Hounsfield units and the mean difference of the reconstructed HU between the planning CT and CBCTs of the virtual human phantom.
RESULTS: For the head and neck phantom, the CNR increased with more advanced scatter correction algorithm and the ASG selectivity, reaching 3.9, 3.7, 3.5 and 3.1 for the high, medium, light and with no grid configuration respectively combined with the iterative software correction. The same is true for the pelvis phantom with CNR improvement reaching 1.5 for the heavy and medium grid, 1.3 for the light grid and 1.1 on its own. The HU RMSD for the head and neck phantom was 22 HU, 13 HU, 12 HU, 6 HU for iterative correction without the grid, with the light grid, medium grid and the heavy grid, respectively. For same correction strategies the values for the pelvis phantom where 170, 120, 34 and 27 HU. The average difference with the PCT of the virtual human phantom were 59 ± 48 HU and 63 ± 59 HU with scans reconstructed with the iterative correction and two higher selectivity grids. Visual inspection revealed similar trends for a head-and-neck and prostate cancer patient.
CONCLUSIONS: The best scatter mitigation strategy was found to be a combination of a grid with selectivity larger than 9, combined with iterative scatter estimation. None of the investigated grids required increasing the imaging dose. The PSF determined using proposed method leads to image quality improvements results for all but one of the investigated scenarios. This article is protected by copyright. All rights reserved.

PMID: 28556204 [PubMed - as supplied by publisher]



http://ift.tt/2rTjYm6

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου

Αρχειοθήκη ιστολογίου