Publication date: Available online 30 July 2018
Source: European Annals of Otorhinolaryngology, Head and Neck Diseases
Author(s): F. Leclercq, C. Renard, C. Vincent
Abstract
Introduction
Understanding speech in noise is a major challenge for most hearing-impaired subjects, with or without hearing aids. To overcome the weaknesses of French-language speech-in-noise tests, we developed a new instrument, with a balanced mix of difficulty of the speech material.
Material and methods
The speech material comprised 127 sentences taken from the "Marginal Benefit from Acoustic Amplification" (MBAA) corpus, each including 3 keywords. The noise was created using the "onde vocale globale" (global vocal wave: OVG), described by Dodelé. The 127 speech/noise pairs were recorded individually after root-mean-square equalization. The first experiment, on 10 normal-hearing adults, determined the signal-to-noise ratio (SNR) associated with 50% correct keyword identification in each sentence (SNR-50), using an ascending method with noise level set at 73 dB SPL. Relative levels between sentences and noise were then adjusted sentence by sentence to achieve an SNR-50 of 0 dB. The second experiment, with 12 normal-hearing adults, validated the equalization of sentence difficulty.
Results
Mean SNR-50 was −6.64 dB (σ = 1.47). Mean adjusted SNR-50 was 0.08 dB (σ = 0.55). Mean psychometric curve slope was 19.3%/dB, with low standard deviations, testifying to the sensitivity of the speech material.
Conclusion
The VRB (vocale rapide dans le bruit: rapid speech in noise) test is based on sentences from the MBAA corpus with background noise based on the OVG at different signal-to-noise ratios. The test is feasible and able to detect slight variations in speech-in-noise performance between subjects.
https://ift.tt/2AxkCeK
Δεν υπάρχουν σχόλια:
Δημοσίευση σχολίου