Publication date: Available online 27 September 2018
Source: Magnetic Resonance Imaging
Author(s): Behzad Mohebbi, Jan Claussen, Bernhard Blümich
Abstract
Single-sided NMR with the NMR-MOUSE is employed for the characterization of fluids in fibrous and open foam materials. One of the key aspects of this study is the quantification of the fluid amount. To this end critical information was provided by a relaxation study. Using 2 mM/L of a Gd3+ relaxation agent the repetition time could be shortened to 250 ms, improving the correlation coefficient between liquid amount and signal amplitude from R2 = 0.893 to R2 = 0.982. To assess reproducibility and instrument precision, calibration experiments were repeated several times and their variation investigated. The results showed that the device is highly precise and robust with a standard deviation for liquid quantification of less than 1%.