Σφακιανάκης Αλέξανδρος
ΩτοΡινοΛαρυγγολόγος
Αναπαύσεως 5 Άγιος Νικόλαος
Κρήτη 72100
00302841026182
00306932607174
alsfakia@gmail.com

Αρχειοθήκη ιστολογίου

! # Ola via Alexandros G.Sfakianakis on Inoreader

Η λίστα ιστολογίων μου

Τετάρτη 3 Μαΐου 2017

Effect of maximum inhalation flow and inhaled volume on formoterol drug deposition in-vitro from an Easyhaler® dry powder inhaler

Publication date: 15 June 2017
Source:European Journal of Pharmaceutical Sciences, Volume 104
Author(s): Mohamad Abadelah, Firisteh Hazim, Henry Chrystyn, Golshan Bagherisadeghi, Hassan Rahmoune, Hassan Larhrib
Most patients using dry powder inhalers (DPIs) do not achieve the inhalation parameters recommended for pharmacopoeial in-vitro dose emission testing. The dose emission characteristics of formoterol from an Easyhaler® have been measured using the Andersen Cascade Impactor (ACI) with a maximum inhalation flow (MIF) of 28.3, 60 and 90L/min and inhaled volumes (Vin) of 240, 750, 1500 and 2000mL. The total emitted dose (TED) was significantly higher at 90L/min (p<0.05), but the difference in the TED between low (28.3L/min) and high (90L/min) flow rate was significantly reduced by increasing the Vin. The fine particle dose (FPD) was higher (p<0.05) at 90L/min for all Vin values compared to 28.3 and 60L/min. Similarly the mass median aerodynamic diameter (MMAD) was smaller at 90L/min across all Vin values. Dose emission characteristics were lower at 240mL for both MIFs. The results for 240mL could be due to an insufficient Vin pulled through the ACI or incomplete emptying of the dose metering cup. This study shows that the FPD, %FPF and MMAD were not significantly affected by the vin≥750mL and that an inhaled volume as low as 750mL could be used with the ACI.

Graphical abstract

image


http://ift.tt/2pzq1tS

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου

Αρχειοθήκη ιστολογίου