Publication date: 15 May 2017
Source:Talanta, Volume 167
Author(s): Tibor Casian, Andra Reznek, Andreea Loredana Vonica-Gligor, Jeroen Van Renterghem, Thomas De Beer, Ioan Tomuță
The objective of this study was to develop, validate and compare NIR and Raman spectroscopic methods for fast characterization in terms of API content and tensile strength of fixed-dose combination tablets containing amlodipine and valsartan.For the APIs assay NIR-transmittance and Raman-reflectance methods were considered, whereas for the tensile strength assay Raman spectra were recorded in reflectance configuration and NIR spectra were recorded in both reflectance and transmittance. Multivariate calibration models (PLS) were built by applying different pre-processing methods (SNV, MSC, SD+SNV) on certain spectral regions. Correlating pre-processed spectral data with tablet properties resulted in highly predictive models except in the case of NIR-transmittance spectra for tensile strength estimation. The best models selected by cross-validation were further validated on independent samples in terms of linearity, trueness, accuracy and precision. Using Bland and Altman analysis the analytical performance of the NIR and Raman methods were compared, demonstrating their similarity considering the investigated applications. The two spectroscopic methods can be used in association to confirm each others results for at-line characterization of the pharmaceutical product.