Use of remote control in the intraoperative telemetry of cochlear implant: multicentric study

Publication date: Available online 18 May 2018

Source: Brazilian Journal of Otorhinolaryngology

Author(s): Liege Franzini Tanamati, Lilian Ferreira Muniz, Paola Angélica Samuel, Maria Valéria Schmidt Goffi-Gomez, Gislaine Richter Minhoto Wiemes, Daniele Penna Lima, Sílvia Badur Curi, Lucia Cristina Onuki, Carla Fortunato Queiroz, Ana Karla Bigois Capistrano, Adriane Lima Mortari Moreti, Márcia Yuri Tsumura Kimura, Valeria Oyanguren, Herbert Mauch

Abstract
Introduction

The conventional evaluation of neural telemetry and impedance requires the use of the computer coupled to an interface, with software that provides visualization of the stimulus and response. Recently, a remote control (CR220®) was launched in the market, that allows the performance of intraoperative tests with minimal instrumentation.

Objective

To evaluate the agreement of the impedance values and neural telemetry thresholds, and the time of performance in the conventional procedure and by the remote control.

Methods

Multicentric prospective cross-sectional study. Intraoperative evaluations of cochlear implants compatible with the use of CR220® were included. The tests were carried out in the 22 electrodes to compare the time of performance in the two situations. The agreement of the neural telemetry threshold values obtained from five electrodes was analyzed, and the agreement of impedance was evaluated by the number of electrodes with altered values in each procedure.

Results

There were no significant difference between the impedance values. There was a moderate to strong correlation between the electrically-evoked compound action potential thresholds. The mean time to perform the procedures using the CR220 was significantly lower than that with the conventional procedure.

Conclusion

The use of the CR220 provided successful records for impedance telemetry and automatic neural response telemetry.

Resumo
Introdução

A avaliação convencional da telemetria neural e de impedâncias implica no uso do computador acoplado a uma interface, com software fornecendo o estímulo e a visualização das respostas. Recentemente, foi lançado um controle remoto (CR220®), que possibilita a realização dos testes intraoperatórios com instrumental mínimo.

Objetivo

Avaliar a concordância dos valores das impedâncias e dos limiares da telemetria neural, e o tempo de execução no procedimento convencional e pelo controle remoto.

Método

Estudo prospectivo transversal multicêntrico. Foram incluídas as avaliações intraoperatórias de implante coclear compatível com o uso do CR220®. Os testes foram realizados nos 22 eletrodos para comparar os tempos de execução nas duas situações. Foi analisada a concordância dos valores do limiar da telemetria neural obtidos em cinco eletrodos, e a concordância das impedâncias foi avaliada pelo número de eletrodos com valores alterados em cada procedimento.

Resultados

Não houve diferença significante entre as impedâncias. Obteve-se moderada a forte correlação entre os limiares do potencial de ação composto eletricamente evocado. O tempo médio para realização dos procedimentos utilizando o CR220 foi significativamente menor que com o procedimento convencional.

Conclusão

O uso do CR220 proporcionou registros bem-sucedidos para a telemetria de impedância e a telemetria automática de respostas neurais.



https://ift.tt/2MRgB64